Allergic Reaction following Implantation of a Nitinol Alloy Inferior Vena Cava Filter.
    Journal of vascular and interventional radiology : JVIR, 2015, Volume: 26, Issue:9

    A 67-year-old woman developed severe pruritus after implantation of a retrievable nitinol inferior vena cava (IVC) filter (55.4% nickel and 44.6% titanium). The pruritus resolved only after filter retrieval. The patient's hypersensitivity to nickel was confirmed by a positive skin patch test substantiating a systemic allergic reaction to the implanted nitinol IVC filter.

    Topics: Aged; Alloys; Dermatitis, Contact; Device Removal; Female; Humans; Prosthesis Implantation; Pruritus; Treatment Outcome; Vena Cava Filters