menaquinone-6 has been researched along with peroxyoxalate* in 1 studies
1 other study(ies) available for menaquinone-6 and peroxyoxalate
Article | Year |
---|---|
Determination of vitamin K homologues by high-performance liquid chromatography with on-line photoreactor and peroxyoxalate chemiluminescence detection.
A sensitive and highly selective high-performance liquid chromatography (HPLC) method was developed for the determination of vitamin K homologues including phylloquinone (PK), menaquinone-4 (MK-4) and menaquinone-7 (MK-7) in human plasma using post-column peroxyoxalate chemiluminescence (PO-CL) detection following on-line ultraviolet (UV) irradiation. The method was based on ultraviolet irradiation (254 nm, 15 W) of vitamin K to produce hydrogen peroxide and a fluorescent product at the same time, which can be determined with PO-CL detection. The separation of vitamin K by HPLC was accomplished isocratically on an ODS column within 35 min. The method involves the use of 2-methyl-3-pentadecyl-1,4-naphthoquinone as an internal standard. The detection limits (signal-to-noise ratio = 3) were 32, 38 and 85 fmol for PK, MK-4 and MK-7, respectively. The recoveries of PK, MK-4 and MK-7 were greater than 82% and the inter- and intra-assay R.S.D. values were 1.9-5.4%. The sensitivity and selectivity of this method were sufficient for clinical and nutritional applications. Topics: Chromatography, High Pressure Liquid; Female; Humans; Luminescence; Male; Oxalates; Reproducibility of Results; Ultraviolet Rays; Vitamin K 1; Vitamin K 2; Vitamins | 2007 |