Application of analytical quality by design principles for the determination of alkyl p-toluenesulfonates impurities in Aprepitant by HPLC. Validation using total-error concept.
    Journal of pharmaceutical and biomedical analysis, 2018, Feb-20, Volume: 150

    In the research presented we report the development of a simple and robust liquid chromatographic method for the quantification of two genotoxic alkyl sulphonate impurities (namely methyl p-toluenesulfonate and isopropyl p-toluenesulfonate) in Aprepitant API substances using the Analytical Quality by Design (AQbD) approach. Following the steps of AQbD protocol, the selected critical method attributes (CMAs) were the separation criterions between the critical peak pairs, the analysis time and the peak efficiencies of the analytes. The critical method parameters (CMPs) included the flow rate, the gradient slope and the acetonitrile content at the first step of the gradient elution program. Multivariate experimental designs namely Plackett-Burman and Box-Behnken designs were conducted sequentially for factor screening and optimization of the method parameters. The optimal separation conditions were estimated using the desirability function. The method was fully validated in the range of 10-200% of the target concentration limit of the analytes using the "total error" approach. Accuracy profiles - a graphical decision making tool - were constructed using the results of the validation procedures. The β-expectation tolerance intervals did not exceed the acceptance criteria of±10%, meaning that 95% of future results will be included in the defined bias limits. The relative bias ranged between - 1.3-3.8% for both analytes, while the RSD values for repeatability and intermediate precision were less than 1.9% in all cases. The achieved limit of detection (LOD) and the limit of quantification (LOQ) were adequate for the specific purpose and found to be 0.02% (corresponding to 48μgg

    Topics: Aprepitant; Benzenesulfonates; Calibration; Chromatography, High Pressure Liquid; Drug Contamination; Limit of Detection; Morpholines; Multivariate Analysis; Quality Control; Reference Standards; Reproducibility of Results; Technology, Pharmaceutical

2018